Type: Error/Fault Category: General Tags: ReplyTo: 62 Subject: Re: Data type change: int8 -> float16 ====== Re: Data type change: int8 -> float16 ====== The TeleDyne LeCroy Control Class (https://github.com/BongHwi/TeleDyneLeCroyControl) was handling the scope data with **8bit type**. This issue has been solved by adding this line: `self.write("CFMT DEF9,WORD,BIN")` for configuring the scope. The attached plot shows the two different waveform shapes (RED: new 12bit, BLUE: old, 8bit) --- Details: The issue has been found by Satoshi, observing the raw wave form with unnatural (descreted signal shape) -> Disussed for the proper shape of the signal -> The current oscilloscope was 12bit type (HD), so it should handle 12bit data (4096 points) -> The current scope class was based on the 8bit data (int8) -> Forcing 16 bit data type (`WORD` instead of `BYTE`) soved the issue. -> Git PR https://github.com/BongHwi/TeleDyneLeCroyControl/commit/ddce5bc11b95bd501fe7f39d3b570cd8c1377d9a